| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Dentomaxillofacial Radiology, Vol 21, Issue 2 88-92, Copyright © 1992 by British Institute of Radiology
ARTICLES |
K. Araki, H. Kitamori, K. Yoshiura, H. Okuda and M. Ohki
Nagasaki University School of Dentistry.
A special apparatus was developed for standardization of the lateral oblique projection of the mandible (LOPM) in order to apply the digital subtraction method. Geometric reproducibility was examined over an 8-week period using a dry skull. Angulation errors were determined by the relation to image noise in the subtraction image. It was found that the relation between image noise and angulation error was approximated by a second-order polynomial. Geometric reproducibility was within 1 degree over the experimental period. These results suggest that this standardized LOPM is satisfactory for clinical application of digital subtraction. Two cases are presented where this technique was used to demonstrate longitudinal changes in bone density.
This article has been cited by other articles:
![]() |
J. Ruijter, J. Verhoeven, J. van der Linden, M. Cune, and M Terlou Image processing and analysis program for measurement of bone density changes in reference and follow-up standardized extraoral oblique lateral cephalometric radiographs of the mandible Dentomaxillofac. Radiol., November 1, 2003; 32(6): 379 - 384. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |