Dentomaxillofacial Radiology, Vol 27, Issue 3 131-135, Copyright © 1998 by British Institute of Radiology
ARTICLES |
K. Yoshiura, H. Stamatakis, X. Q. Shi, U. Welander, W. D. McDavid, J. Kristoffersen and G. Tronje
Department of Oral Radiology, Karolinska Institutet, Stockholm, Sweden.
OBJECTIVE: To compare the effect of an additional scintillator layer on the psychophysical properties of a CCD detector for digital dental radiography. METHODS: Radiographs of a test object containing ten holes of increasing depth were obtained throughout the exposure range of two CCD detectors at 70 and 90 kVp. One was the original detector for the Sens-A-Ray system (Regam Medical Systems, Sundsvall, Sweden) and the other the same detector covered by a scintillator layer. Ten viewers evaluated the radiographs for the number of perceptible holes. From these data and the dose response functions for the two detectors, the minimum perceptible exposure difference was found. The reciprocal values of this parameter were plotted against the logarithm of exposure to create Perceptibility Curves (PCs). RESULTS: The four PCs had essentially the same shape and height. There was a shift to lower exposures in the PCs for the detector covered by a scintillator. CONCLUSIONS: The two detectors have essentially the same psychophysical properties. Since the detector covered by a scintillator layer is more sensitive, it should be preferred for clinical practice since the dose to the patient is reduced.
This article has been cited by other articles:
![]() |
X-Q Shi, G Li, K Yoshiura, and U Welander Perceptibility curve test for conventional and colour-coded radiographs Dentomaxillofac. Radiol., September 1, 2004; 33(5): 318 - 322. [Abstract] [Full Text] [PDF] |
||||
![]() |
M Mastoris, K Yoshiura, U Welander, K Tsiklakis, E Papadakis, and G Li Psychophysical properties of a new F-speed intraoral film Dentomaxillofac. Radiol., May 1, 2004; 33(3): 158 - 163. [Abstract] [Full Text] [PDF] |
||||
![]() |
K Yoshiura, U Welander, W. McDavid, G Li, X-Q Shi, E Nakayama, M Shimizu, K Okamura, and S Kanda Comparison of the psychophysical properties of various intraoral film and digital systems by means of the perceptibility curve test Dentomaxillofac. Radiol., March 1, 2004; 33(2): 98 - 102. [Abstract] [Full Text] [PDF] |
||||
![]() |
G Li, U Welander, K Yoshiura, X-Q Shi, and W. McDavid Perceptibility curve test for digital radiographs before and after correction for attenuation and correction for attenuation and visual response Dentomaxillofac. Radiol., November 1, 2003; 32(6): 372 - 378. [Abstract] [Full Text] [PDF] |
||||
![]() |
K Yoshiura, U Welander, W. McDavid, G Li, X-Q Shi, T Kawazu, M Tatsumi, T Matsuo, and S Kanda Perceptibility curves for the Digora system Dentomaxillofac. Radiol., May 1, 2003; 32(3): 191 - 197. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |