Dentomaxillofacial Radiology, Vol 27, Issue 6 344-350, Copyright © 1998 by British Institute of Radiology
ARTICLES |
M. G. Cavalcanti and M. W. Vannier
Department of Radiology, Faculty of Odontology, University of Sao Paulo, Brazil.
OBJECTIVE: To determine the relative accuracy of craniofacial measurements obtained from volumetric spiral computed tomography (CT) in two-dimensional (2D) and three-dimensional (3D) reconstructions. MATERIALS AND METHODS: Nine cadaver heads were examined with spiral CT at 3 mm collimation with 2 mm/s table feed. The CT data, stored on optical disks to allow full retrospective review of any image, were transferred to a networked computer workstation for off-line visualization, measurement and analysis. Three-dimensional surface images (3DCT) were generated from spiral CT scans (2DCT). Measurements (n = 28) were made from (both unique and conventional craniometric anatomical) landmarks (n = 23) identified in 2DCT and 3DCT images by two observers twice. The soft tissues were then removed from the cadavers and the measurements performed with an electromagnetic digitizer. RESULTS: The accuracy for 3DCT was higher than 2DCT images for seven of 28 measurements. Twenty five per cent of the 2DCT measurements were found to be significantly different from the physical measurements. All measurements with 3DCT were found to have satisfactory accuracy in comparison with physical measurements. CONCLUSIONS: Three-dimensional spiral CT can distinguish craniofacial anatomy more accurately with some measurements.
This article has been cited by other articles:
![]() |
S. Connor, T Arscott, J Berry, L Greene, and R O'Gorman Precision and accuracy of low-dose CT protocols in the evaluation of skull landmarks Dentomaxillofac. Radiol., July 1, 2007; 36(5): 270 - 276. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Draenert, E Coppenrath, P Herzog, S Muller, and U. Mueller-Lisse Beam hardening artefacts occur in dental implant scans with the NewTom(R) cone beam CT but not with the dental 4-row multidetector CT Dentomaxillofac. Radiol., May 1, 2007; 36(4): 198 - 203. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Lascala, J Panella, and M. Marques Analysis of the accuracy of linear measurements obtained by cone beam computed tomography (CBCT-NewTom) Dentomaxillofac. Radiol., September 1, 2004; 33(5): 291 - 294. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Cavalcanti, S. Rocha, and M. Vannier Craniofacial measurements based on 3D-CT volume rendering: implications for clinical applications Dentomaxillofac. Radiol., May 1, 2004; 33(3): 170 - 176. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |