Dentomaxillofacial Radiology, Vol 28, Issue 2 80-88, Copyright © 1999 by British Institute of Radiology


ARTICLES

Effects of developer exhaustion on the sensitometric properties of four dental films

K. Syriopoulos, X. L. Velders, G. C. Sanderink, F. C. van Ginkel and P. F. van der Stelt
Department of Oral Radiology, Academic Centre for Dentistry Amsterdam (ACTA), The Netherlands.

OBJECTIVES: To examine the effects of exhaustion of five different processing solutions on the sensitometric properties of four dental X-ray films: Ektaspeed Plus and Ultra-speed (Kodak Eastman Co. Rochester, USA) and new and previous Dentus M2 (Agfa-Gevaert, Mortsel, Belgium). METHODS: An aluminum stepwedge was used to construct characteristic curves for the four films. All films were processed manually using three sets of chemicals for manual processing: Agfa (Heraeus Kulzer, Dormagen, Germany), Kodak (Kodak-Pathe, Paris, France) and Demat (Viscopac, Athens, Greece) and two sets for automatic processing: Durr XR and Periomat (Durr Dental, Bietigheim-Bissingen, Germany). Film speed and gradient were evaluated until the chemicals were completely exhausted. An analysis of variance was performed separately for each set of chemicals for manual and automatic processing. RESULTS: Ektaspeed Plus was the fastest film in the manual processing solutions. The new Dentus M2 and Ektaspeed Plus films had similar speed using the chemicals for automatic processing. Ultra-speed had the lowest speed in all solutions, but it had the greatest consistency. Exhaustion of the developer caused a comparable decrease in speed of Ektaspeed Plus and the two Dentus M2 films. In fresh chemistry Agfa was the strongest manual processing solution, but it had the highest exhaustion rate. The Durr XR chemical was stronger than Periomat over the whole test period. CONCLUSIONS: The combination of film and processing solution is an important factor for achieving constant sensitometric properties. Ektaspeed Plus and the new Dentus M2 film should be used in dental practice, as they require lower exposure and have equivalent or superior properties compared with Ultra-speed.


This article has been cited by other articles:


Home page
Dentomaxillofac RadiolHome page
M. Casanova and F Haiter-Neto
Effects of developer depletion on image quality of Kodak Insight and Ektaspeed Plus films
Dentomaxillofac. Radiol., March 1, 2004; 33(2): 108 - 113.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING ALL BIR JOURNALS
Copyright © 1999 by the British Institute of Radiology.