DMFR
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hayakawa, Y.
Right arrow Articles by Nishizawa, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hayakawa, Y.
Right arrow Articles by Nishizawa, K.

Dentomaxillofacial Radiology, Vol 30, Issue 5 285-292, Copyright © 2001 by British Institute of Radiology


ARTICLES

Paediatric absorbed doses from rotational panoramic radiography

Y. Hayakawa, N. Kobayashi, K. Kuroyanagi and K. Nishizawa
Department of Oral and Maxillofacial Radiology, Tokyo Dental College, Chiba, Japan.

OBJECTIVES: To determine the paediatric doses in rotational panoramic radiography with film/screen and photostimulable phosphor receptors. STUDY DESIGN: A paediatric anthropomorphic head and neck phantom was used. Absorbed doses were measured for two panoramic systems, the Orthophos (Sirona Dental Systems, Bensheim, Germany) and the PM 2002 CC (Planmeca Oy, Helsinki, Finland), with and without programmable child settings, using both screen/film and photostimulable phosphor receptors. RESULTS: Absorbed doses to the eye ranged from 5 to 24 micro Gy. Doses to the dental arches with the Orthophos unit ranged from 50 to 555 micro Gy with the adult and from 27 to 436 micro Gy with the child program; using the PM 2002 CC unit, doses ranged from 56 to 1040 micro Gy using the adult settings, and from 60 to 890 micro Gy with the paediatric settings. The paediatric exposure settings reduced doses at most locations for both panoramic systems. The highest doses were measured near the rotational axes of the X-ray beam. Paediatric settings with the Orthophos P10 resulted in the dose reduction more than 50% to the thyroid but not with the PM 2002 CC. When lower kVcp or mA settings were used, absorbed doses were effectively reduced for all combinations of machines, programs and detectors. CONCLUSIONS: Specific program settings for children reduced the absorbed doses from panoramic radiography irrespective of the machine or receptor used.