DMFR
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yoshiura, K
Right arrow Articles by Kanda, S
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yoshiura, K
Right arrow Articles by Kanda, S
Dentomaxillofacial Radiology (2003) 32, 191-197
© 2003 British Institute of Radiology
doi: 10.1259/dmfr/25806978


TECHNICAL REPORT

Perceptibility curves for the Digora system

K Yoshiura*,1,2, U Welander2, WD McDavid3, G Li2, X-Q Shi2, T Kawazu1, M Tatsumi1, T Matsuo1 and S Kanda1

1 Department of Oral and Maxillofacial Radiology, Faculty of Dental Science, Kyushu University, Fukuoka, Japan; 2 Department of Oral Radiology, Karolinska Institutet, Stockholm, Sweden; 3 Department of Dental Diagnostic Science, University of Texas Health Science Center at San Antonio, San Antonio, Texas, USA

*Correspondence to: Kazunori Yoshiura, Department of Oral and Maxillofacial Radiology, Faculty of Dental Science, Kyushu University, Maidashi 3-1-1, Higashi-ku, Fukuoka, 812-8582 Japan; Email: yoshiura{at}rad.dent.kyushu-u.ac.jp

Received 26 February 2003; revised 30 May 2003; accepted 5 June 2003

Objectives: To construct perceptibility curves (PCs) for given calibration settings in order to define psychophysical properties of the Digora storage phosphor system and to evaluate the effects of automatic exposure correction (AEC) on the PCs.

Methods: The Digora system was calibrated at two exposures, 80 µC kg-1 (high calibration) and 40 µC kg-1 (low calibration). Since the grey levels displayed in the radiographs are adjusted by AEC, dose–response functions at high calibration were obtained using AEC on and off modes. The dose–response function at low calibration was obtained with AEC off. The PC at each experimental setting was calculated using known physical parameters of the system and the performance of the average observer used in a previous study. In addition, PCs were also constructed using transmitted radiation flux behind the test object calculated from the attenuation coefficient in order to study observer and system performance. PCs obtained under these conditions were compared.

Results: The PC using calculated transmitted radiation flux behind the test object showed a wide plateau at the peak owing to AEC, while the PC obtained by a modified approach showed a higher but narrower peak. There were no differences between the two PCs using AEC on and off modes when the PCs were constructed using a modified approach. There were no differences between the two PCs obtained at high and low calibration settings or between the three PCs obtained with AEC on except for the position along the exposure axis.

Conclusions: Psychophysical properties of the Digora system may be determined if we employ registered exposures from a dose–response function with AEC off under a given calibration setting. Under these circumstances the shape and height of the PCs will be unchanged irrespective of the AEC mode.

Keywords: radiography, dental; digital radiography, dental; perception




This article has been cited by other articles:


Home page
Dentomaxillofac RadiolHome page
K Yoshiura, U Welander, W. McDavid, G Li, X-Q Shi, E Nakayama, M Shimizu, K Okamura, and S Kanda
Comparison of the psychophysical properties of various intraoral film and digital systems by means of the perceptibility curve test
Dentomaxillofac. Radiol., March 1, 2004; 33(2): 98 - 102.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING ALL BIR JOURNALS
Copyright © 2003 by the British Institute of Radiology.