DMFR
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bahrami, G
Right arrow Articles by Wenzel, A
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bahrami, G
Right arrow Articles by Wenzel, A
Dentomaxillofacial Radiology (2003) 32, 317-321
© 2003 British Institute of Radiology
doi: 10.1259/dmfr/14212871


RESEARCH

Bitewing examination with four digital receptors

G Bahrami*, C Hagstrøm and A Wenzel

Department of Oral Radiology, Royal Dental College, University of Aarhus, Vennelyst Boulevard, 8000 Aarhus C, Denmark

*Correspondence to: Golnosh Bahrami, DDS, Department of Oral Radiology, Royal Dental College, Vennelyst Boulevard, DK - 8000 Aarhus C, Denmark; Email: gbahrami{at}odont.au.dk

Received 28 January 2003; revised 28 March 2003; accepted 14 September 2003

Objectives: To evaluate recording errors and patient discomfort during bitewing examinations using four digital receptors.

Methods: Seventy-eight patients had two bitewings taken on each side of the mouth with the intention of displaying the tooth surfaces from the distal surface of the canine to the distal surface of the most posterior molar, using four digital receptors, two charge-coupled device (CCD) and two photostimulable phosphor (PSP) systems. The patients scored their feelings of discomfort during the examination on a visual analogue scale. Receptor positioning errors in the sagittal plane were determined from the tooth surfaces present on each image and in the vertical plane from the presence of the alveolar bone crest. Cone positioning errors were determined from cone cuts.

Results: Canine and premolar surfaces were more often not depicted on the CCD images than on the PSP images (P<0.05). Cone cuts occurred in 19% of DenOptix images, in 9% of Digora images and in one Planmeca image. The bone crest was more often missing in the upper jaw on Planmeca images than on PSP images (P<0.01). In the lower jaw, Trophy images more often missed the bone crest than the other systems (P<0.05). Patients ranked the receptors as follows (with increasing discomfort): DenOptix, Planmeca, Digora and Trophy, with all being significantly different (P<0.05).

Conclusions: It was more difficult to correctly position CCD sensors than PSP plates in the vertical plane, resulting in more images with missing alveolar bone crest. CCD sensors most often did not display the most anterior surfaces in a bitewing examination.

Keywords: radiography, digital; radiography, bitewing




This article has been cited by other articles:


Home page
Journal of the American Dental AssociationHome page
F. Haiter-Neto, A. dos Anjos Pontual, M. Frydenberg, and A. Wenzel
A comparison of older and newer versions of intraoral digital radiography systems: Diagnosing noncavitated proximal carious lesions
J Am Dent Assoc, October 1, 2007; 138(10): 1353 - 1359.
[Abstract] [Full Text] [PDF]


Home page
Dentomaxillofac RadiolHome page
J. Brian and G. Williamson
Digital radiography in dentistry: a survey of Indiana dentists
Dentomaxillofac. Radiol., January 1, 2007; 36(1): 18 - 23.
[Abstract] [Full Text] [PDF]


Home page
Dentomaxillofac RadiolHome page
A Wenzel
A review of dentists' use of digital radiography and caries diagnosis with digital systems.
Dentomaxillofac. Radiol., September 1, 2006; 35(5): 307 - 314.
[Abstract] [Full Text] [PDF]


Home page
J. Dent. Res.Home page
A. Wenzel
Bitewing and Digital Bitewing Radiography for Detection of Caries Lesions
J. Dent. Res., July 1, 2004; 83(suppl_1): C72 - C75.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING ALL BIR JOURNALS
Copyright © 2003 by the British Institute of Radiology.