| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
RESEARCH |
1 Department of Oral Surgery, Faculty of Dentistry, Heinrich Heine University Düsseldorf, Moorenstraße 5, 40225 Düsseldorf, Germany; 2 Department for Computer Science, Heinrich Heine University Düsseldorf, Moorenstraße 5, 40225 Düsseldorf, Germany
*Correspondence to: Dr Andreas Künzel, Poliklinik für Zahnärztliche Chirurgie und Aufnahme, Moorenstraße 5, 40225 Düsseldorf, Germany; Email: kuenzel{at}uni-duesseldorf.de
Received 8 November 2002; revised 18 October 2003; accepted 10 December 2003
Objectives: The visually utilizable resolution of intraoral dental films was examined with the naked eye, with two-fold optical magnification and after digitization using a computer display.
Methods: Agfa Dentus M2®, Kodak Ektaspeed Plus® and Kodak InSight® dental films were exposed with a line pair (lp) test pattern providing frequencies up to 16.6 lp mm1. Films were developed at 24°C in a Dürr Periomat machine and at 28°C in a Dürr AC245 machine. Forty dental students evaluated the maximum visually detectable resolution with their naked eye, with an X-ray viewer® (providing two-fold magnifying lenses and eliminating disturbing light) and following digitization and monitor display examination.
Results: The best detectable resolution was achieved through digitization, reaching a mean of up to 16.5 lp mm1. With two-fold magnification, a mean of up to 13.3 lp mm1 could be resolved, while only 11.7 lp mm1could be resolved with the naked eye. Unlike the other film types, the resolution of Ektaspeed depended on the processing type when viewing digitized images or when viewing with the naked eye. There was a trend for students above 29 years of age to detect a lower resolution with the naked eye, although they performed comparably with students below 29 years when using the X-ray viewer® or viewing digitized images.
Conclusions: Film resolution is utilized best through digitization and secondly using a magnifying lens. With the naked eye, a mean of 11 lp mm1 with a broad distribution can be resolved. A magnifying lens is recommended if resolution is important or if the viewer's eyesight is reduced. Compared with E-speed films, Kodak Insight was less dependent on processing conditions.
Keywords: radiology; digitization; line pairs; observer performance; film processing
This article has been cited by other articles:
![]() |
N Parissis, A Kondylidou-Sidira, A Tsirlis, and P Patias Conventional radiographs vs digitized radiographs: image quality assessment Dentomaxillofac. Radiol., November 1, 2005; 34(6): 353 - 356. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. B Ludlow and A. Mol Resolution of digitized intraoral dental films Dentomaxillofac. Radiol., May 1, 2004; 33(3): 208 - 208. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |