DMFR
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Shi, X-Q
Right arrow Articles by Welander, U
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Shi, X-Q
Right arrow Articles by Welander, U
Dentomaxillofacial Radiology (2004) 33, 318-322
© 2004 British Institute of Radiology
doi: 10.1259/dmfr/27372105


RESEARCH

Perceptibility curve test for conventional and colour-coded radiographs

X-Q Shi*,1,2, G Li1, K Yoshiura1 and U Welander1

1 Department of Oral Radiology, Institute of Odontology, Karolinska Institutet, Stockholm, Sweden; 2 Department of Cariology and Endodontology, Institute of Odontology, Karolinska Institutet, Stockholm, Sweden

*Correspondence to: Xie-Qi Shi, Department of Cariology and Endodontology, Institute of Odontology, Karolinska Institutet, Box 4064, S-141 04 Huddinge, Sweden; Email: xie.qi.shi{at}ofa.ki.se

Received 23 February 2004; accepted 11 August 2004

Objective: A colour scale has been developed as a complement to the grey scale in radiographs. The aim of the present study was to compare colour-coded with conventional black-and-white radiographs in terms of the perceptibility curve test.

Material and methods: Twelve digital radiographs were exposed on a test object from low to high exposures and recorded using the DIXI system (Planmeca Oy, Helsinki, Finland). An aluminium test object was used containing ten object details in the form of holes with depths ranging from 0.03 mm to 0.30 mm in steps of 0.03±0.01 mm. The new colour scale was used to transform the conventional black-and-white radiographs into colour radiographs by specially designed software. Ten observers were asked to analyse four sets of radiographs, i.e. colour-coded and black-and-white radiographs, and reversed colour-coded and reversed black-and-white radiographs. The object detail having the lowest perceptible contrast in each radiograph was registered for each observer. Perceptibility curves were plotted based on the mean value of observer data.

Results: Results from the perceptibility curve test showed that the information in colour-coded radiographs was at least as good as that in black-and-white radiographs. In fact, in the lower exposure range, colour-coded radiographs exhibited better perception than conventional black-and-white radiographs.

Conclusion: Radiographs that are colour-coded with the applied colour scale may be used as an alternative to conventional black-and-white radiographs.

Keywords: radiography, dental; digital radiography, dental; technology, radiographic; perception, colour perception




This article has been cited by other articles:


Home page
Journal of the American Dental AssociationHome page
E. T. Parks
Digital Radiographic Imaging: Is the Dental Practice Ready?
J Am Dent Assoc, April 1, 2008; 139(4): 477 - 481.
[Abstract] [Full Text] [PDF]


Home page
Dentomaxillofac RadiolHome page
E Alpoz, E Sogur, and B. Baksi Akdeniz
Perceptibility curve test for digital radiographs before and after application of various image processing algorithms
Dentomaxillofac. Radiol., December 1, 2007; 36(8): 490 - 494.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING ALL BIR JOURNALS
Copyright © 2004 by the British Institute of Radiology.