| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
RESEARCH |
1 Department of Oral Radiology, Academic Centre for Dentistry Amsterdam (ACTA), Amsterdam, The Netherlands; 2 Division of Oral and Maxillofacial Radiology, Department of Diagnostic Sciences and General Dentistry, University of North Carolina at Chapel Hill, NC, USA
*Correspondence to: AN van Daatselaar, Department of Oral Radiology, Academic Center for Dentistry Amsterdam (ACTA), Louwesweg 1, 1066EA Amsterdam, The Netherlands; Email: a.v.daatselaar{at}acta.nl
Received 25 September 2003; revised 27 August 2004; accepted 31 August 2004
Objectives: To investigate the relationship between the number of basis projections for local computed tomography (CT) and the detection of proximal caries and to find a minimum number of projections needed to maintain diagnostic accuracy.
Methods: We presented observers (n=12) with stacks of both axial and vertical CT slices of 23 extracted teeth placed in a dry human mandible. The slices were prepared with 14, 20, 33 and 100 basis projections. The observers scored the proximal surfaces for the presence of caries on a 15 confidence scale. The performance of the varying number of projections was compared with conventional digital radiographs.
Results: The performance of all four CT modalities was significantly better than conventional radiographs (P=0.005 to P=0.021) and showed a consistent increase with the number of projections. Diagnostic performance depended significantly on lesion depth (P=0.00), but not on observer.
Conclusions: We conclude that the number of CT projections used can be reduced at least to 20 with the diagnostic performance still markedly better than that of conventional film, provided that the observer can make use of both axial and vertical stacks of CT slices.
Keywords: tomography, X-ray computed; radiography, dental, digital; dental caries; observer performance
This article has been cited by other articles:
![]() |
M. Mora, A Mol, D. Tyndall, and E. Rivera Effect of the number of basis images on the detection of longitudinal tooth fractures using local computed tomography Dentomaxillofac. Radiol., October 1, 2007; 36(7): 382 - 386. [Abstract] [Full Text] [PDF] |
||||