Dentomaxillofacial Radiology (2007) 36, 282-284
© 2007 British Institute of Radiology
doi: 10.1259/dmfr/67494525


RESEARCH

Patient radiation dose management in dental facilities according to the X-ray focal distance and the image receptor type

L Gonzalez1,* and J Moro2

1 Department of Radiology, Medicine School, Complutense University, 28040 Madrid, Spain; 2 INFOCITEC, Avda. Monforte de Lemos, 142, 28039 Madrid, Spain

*Correspondence to: Prof. Luciano Gonzalez, Department of Radiology, Medicine School, Complutense University, 28040 Madrid, Spain; Email: luciano{at}med.ucm.es

Received 16 June 2006; revised 15 September 2006; accepted 2 October 2006

Objectives: To determine particular reference values for patient radiation dose management according to the focus-to-skin distance and the image receptor type.

Methods: Air-free ionization chambers and solid state detectors built in quality control devices were used to measure X-ray tube output in a sample of 2811 X-ray units. The mean exposure time was estimated by the local person responsible for the quality assurance programme. From this data, mean air kerma values were calculated at the focus-to-skin distance. To obtain the entrance surface air dose, a backscatter factor of 1.2 was used.

Results: Third-quartile values were 5.3 mGy for old X-ray sets with around 10 cm focus-to-skin distance (FSD); for 20 cm FSD, the values were 4.1 mGy for D type, 3.4 mGy for E/F film type and 1.2 mGy for digital image receptors; and for 30 cm FSD the values were 2.1 mGy for D type, 1.8 mGy for E/F film type and 0.6 mGy for digital image receptors.

Conclusions: Evidence based on third-quartile values different from those expected justifies the need for particular reference values taking into account image system and focal distance, as a way to improve patient radiation dose management.

Keywords: radiation protection; intraoral radiography; patient radiation doses; dose reference values